中国药物警戒 ›› 2017, Vol. 14 ›› Issue (1): 27-31.

• 质量与工艺研究 • 上一篇    下一篇

甲硝唑片仿制药在不同介质中的体外溶出度比较

宋金春1, 侯亚婷2, *   

  1. 1武汉大学人民医院,湖北 武汉 430060;
    2.武汉大学药学院,湖北 武汉 430072
  • 收稿日期:2017-02-16 修回日期:2017-02-16 出版日期:2017-01-20 发布日期:2017-02-16
  • 作者简介:宋金春,男,博士,主任药师,临床药学。

Comparison of Dissolubility in Vitro of Generic Metronidazole Tablets in Different Mediums

SONG Jin-chun1, HOU Ya-ting2, *   

  1. 1Renmin Hospital of Wuhan University, Hubei Wuhan 430060, China;
    2 School of Pharmacy, Wuhan University, Hubei Wuhan 430072, China
  • Received:2017-02-16 Revised:2017-02-16 Online:2017-01-20 Published:2017-02-16
  • Contact: 侯亚婷,女,在读硕士,药剂学。E-mail:948669233@qq.com

摘要: 目的建立甲硝唑片的体外溶出度测定方法,将国内外仿制药的溶出曲线进行比较评价其一致性,考察国内药质量,为临床用药提供参考。方法采用高效液相色谱法测定溶出液的含量,绘制溶出曲线。色谱柱为Agilent TC-C18,流动相为甲醇-水(20:80),流速为1.0 mL·min-1,检测波长为320 nm,柱温为30℃,进样量为20 μL。采用桨法,转速为50 rpm,分别以水、pH=1.2的盐酸溶液、pH=4.0醋酸盐缓冲液、pH=6.8磷酸盐缓冲液为溶出介质测定溶出度,溶出介质体积为900 mL,采用相似因子(f2)评价溶出曲线的相似性。结果所有厂家的溶出度均符合药典规定,国内甲硝唑片仿制药溶出曲线与国外不相似。结论部分国产仿制药质量有待改善,应加强仿制药质量监督,提高药品质量。

关键词: 甲硝唑片, 溶出曲线, HPLC, 一致性评价

Abstract: Objective To establish a method for determining the dissolubility of metronidazole tablets in vitro, compare the dissolution curves between domestic and foreign generics to evaluate the quality consistency, investigate the quality of domestic drugs and make reference for clinical use. Methods The content of solution was determined by HPLC, and the dissolution curves were drawn. The column was Agilent TC-C18, mobile phase was methanol-water (20:80, V/V) with a flow rate of 1 mL·min-1, detection wavelength was 320 nm, column temperature was 30°C, injection volume was 20 μL. The stirring paddle method was adopted, the dissolution mediums were water, pH=1.2 hydrochloric acid solution, pH=4.0 acetic acid-sodium acetate buffer solution, pH=6.8 phosphate buffer solution, the volume of which was 900 mL, the rotation speed was 50 rpm. The similarity of dissolution curve was evaluated by similar factor (f2). Results The dissolution of metronidazole tablets made by all manufacturers were conformed to the requirements of the Chinese Pharmacopoeia, but the dissolution curves of generic drugs from domestic manufacturers were not similar with foreign countries’. Conclusion The quality of some domestic generics needs to be further improved. It is necessary to strengthen the supervision of generics in order to enhance the quality.

Key words: metronidazole tablet, dissolution curve, HPLC, consistency evaluation

中图分类号: